FEATURES
Calibrated and temperature compensated
pressure sensor with 0 .. 5 V output voltage
Differential / gage, bidirectional differential,
absolute and barometric versions
Wide variety of pressure ranges:
5 mbar up to 2 bar
Piezoresistive sensing element
High precision digital signal conditioning
High accuracy at room temperature
Low overall error within a temperature range
of -25 .. +85°C
Supply voltage range: 8 .. 36 V
Compact and robust package
Dust- and water-tight (IP67)
M5 circular sensor connector
Ready to use, simple and quick installation
RoHS and REACH compliant
TYPICAL APPLICATIONS
Static pressure sensing
Dynamic pressure sensing
Barometric pressure sensing
Vacuum measurement
Level measurement
Gas flow measurement
Heating, Ventilation and Air Conditioning
(HVAC)
GENERAL DESCRIPTION
AMS 4711 is a series of miniaturized pressure
transmitters suitable for industrial applications and
high precision measurements. All AMS 4711 can
be powered within a supply voltage range of
VS = 8 .. 36 V and provide an industrial output
voltage range of VOUT = 0 .. 5 V. They are
calibrated and temperature compensated in a wide
temperature range of -25 .. +85°C.
AMS 4711 comes in a robust plastic package and
is ready to use. Pressure is either applied via two
barbed side ports for differential (relative) or
bidirectional differential pressure transmitter types
or via a single port for absolute or barometric
types. The electrical connection is realized with an
M5 circular sensor connector. AMS 4711’s
package is dust- and water-tight conforming to
IP67.
The pressure transmitters in the AMS 4711 series
are available for various applications and pressure
ranges: Differential (relative) devices in pressure
ranges from 0 .. 5 mbar up to 0 .. 2 bar, absolute
pressure variants for 0 .. 1 bar and 0 .. 2 bar as
well as a barometric type for 0.7 .. 1.2 bar.
Bidirectional-differential devices are available from
-5 .. +5 mbar up to -1 ..+1 bar for measurement of
positive and negative differential pressure. Other
pressure ranges and customized output voltage
ranges are available on request
|