22 / 29 page ![]() 22 PROBLEM POTENTIAL CAUSES SOLUTIONS IntelliTune fails because of a “load short” error VLIM may be too high. Reduce the current limit and rerun IntelliTune to automatically reduce VLIM. Then reset the current limit and rerun the scan. Use remote command TEC:VLIM 9 to set the VLIM to 9 V and then rerun IntelliTune. Put a resistor in series with the thermoelectric to increase the load compliance voltage. Load Fault Error is triggered A Load Fault Error may be falsely triggered when operating near maximum current and near maximum voltage. Use remote command TEC:VLIM 27 to limit the internal voltage supply to 27 V instead of the 28 V maximum. |
|
Link URL |
| War ALLDATASHEET hilfreich? [ DONATE ] |
Über Alldatasheet | Werbung | Kontakt | Privatsphäre und Datenschutz | Link zum Datenblatt | Linktausch | Hersteller All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|