5 / 31 page ![]() 5 ADJUST TEMPERATURE SETPOINT 1. Press to access the TC Control Screen. 2. Press the blue Setpoint temperature value and turn the adjustment knob to change it. The instrument will beep when set. The TC Control Screen is also the location of the PID control parameters and IntelliTune (intelligent autotune) characterization scan. The PID values are manually editable or can be adjusted automatically via IntelliTune. ENABLE TC CURRENT Press the TC ENABLE button on the front panel. There is no delay before the button illuminates and the current flows to the thermoelectric. When current to the thermoelectric is enabled, this light will be illuminated blue. Using a 10 kΩ test load, temperature will not stabilize. Current will drive to the positive limit if setpoint temperature is below 25ºC and to the negative limit if setpoint is above 25ºC. SET THE LIMITS – LD 1. Press to access the Laser Diode Driver Control Screen. 2. Access the LD Limit Screen, press the blue text to make it editable, then use the adjustment knob to configure the limits as required. If current reaches the set limit, output current will remain at the limit. Output will not be disabled if the limit is reached. CHOOSE LD CONTROL MODE On the LD Control Screen, choose either Constant Current or Constant Power mode. Pressing the text toggles the selection. It is not editable while LD current is enabled. CONSTANT CURRENT MODE Constant Current (CC) mode maintains a constant current through the laser. The setpoint corresponds to the current through the laser. CONSTANT POWER MODE Constant Power (CP) mode maintains a constant current through the monitor photodiode, corresponding to a constant power output. If CP mode is chosen, additional parameters must be set on the Photodiode Screen. The test load will not work in CP mode. 1. Press to access the Photodiode Screen. 2. Press the blue text next to "PD Biasing" to toggle whether the bias voltage for the photodiode is enabled or disabled. See the Photodiode Basics App Note for more detail. 3. Press the blue text next to "PD Range" until the value matches the appropriate current range for the photodiode. The available options are 500 μA, 5 mA, and 10 mA. 4. If the transfer function relating photodiode current to optical power is known, this value can be entered using the "Transfer Function" setting. Units are W/mA. SET LASER TYPE Access the LD Control Screen. Edit the laser type by pressing the blue text to toggle between Types A/B and Type C lasers. Type A Laser Diode Type B Laser Diode Type C Laser Diode Common Cathode Laser Diode Anode & Photodiode Cathode Common Isolated Photodiode Short the Laser Diode Anode to Photodiode Cathode Common Anode Laser Diode Cathode & Photodiode Anode Common Figure 4. Laser Type Diagrams NOTE: Type A/B lasers use a separate connection on the rear panel from Type C lasers. Ensure that the proper rear panel connection is made in addition to selecting the proper type on the LD Control Screen. SET LASER SETPOINT Also from the LD Control Screen, edit the laser current setpoint by pressing the blue text and turning the adjustment knob. Note: For added safety, if the setpoint is set to the value of the current limit, actual current will be ~43 mA below setpoint. ENABLE LASER CURRENT Press the LDD ENABLE button on the front panel. By default, there is a 2 second delay (user-adjustable via remote commands) between the time the button is pushed and current being supplied to the load. When laser current is enabled, this light will be illuminated solid red. MONITOR ALL I/V OUTPUTS Press to access the Dual Monitor Screen. From here, both the TC and LD output parameters are displayed. Parameters are not editable from this screen. SHUTDOWN THE INSTRUMENT If either of the Enable buttons are illuminated, current is still flowing to the load. Before shutting down, the current needs to be turned off. 1. Press the Enable button. The button light will turn off and current to the test load will stop. 2. On the back panel, press the power switch to the O position. The instrument will shut down. |
|
Link URL |
| War ALLDATASHEET hilfreich? [ DONATE ] |
Über Alldatasheet | Werbung | Kontakt | Privatsphäre und Datenschutz | Link zum Datenblatt | Linktausch | Hersteller All Rights Reserved©Alldatasheet.com |
| Russian : Alldatasheetru.com | Korean : Alldatasheet.co.kr | Spanish : Alldatasheet.es | French : Alldatasheet.fr | Italian : Alldatasheetit.com Portuguese : Alldatasheetpt.com | Polish : Alldatasheet.pl | Vietnamese : Alldatasheet.vn Indian : Alldatasheet.in | Mexican : Alldatasheet.com.mx | British : Alldatasheet.co.uk | New Zealand : Alldatasheet.co.nz |
|
Family Site : ic2ic.com |
icmetro.com |